Description

Background
This solicitation is issued by the U.S. Army for a new Inductively Coupled Plasma - Optical Emission Spectrometry (ICP-OES) system, specifically designed to meet the needs of the agency in analyzing both aqueous and organic samples. The goal of this contract is to acquire a high-performance ICP-OES system that adheres to strict technical specifications and performance requirements, ensuring compliance with EPA detection limits and certified metal standards accuracy.

Work Details
The required instrument is a new, benchtop, dual-view ICP-OES system for simultaneous analysis of aqueous and organic samples. Key specifications include:

1. Spectrometer Requirements:
- Simultaneous reading ICP OES with solid-state detector.
- Bench top design only; floor mount not acceptable.
- Echelle polychromator, 253 mm focal length, single detector, single entrance slit.
- Resolution: less than 0.0065 nm @ 200 nm; less than 0.013 nm @ 327.4 nm.
- Purged + thermostated optical enclosure; purge 0.7 L/min with 3 L/min boost option (less than 189 nm).
- Vertical plasma; single piece torch with optional torches available.
- View equal to or greater than 96% emission spectrum from 167-785 nm.

2. ICP System Requirements:
- Must run aqueous + organic matrices including oil, MIBK, kerosene, hexanes.
- Monitors gas pressures/flows, water flows, air pressure, exhaust, plasma stability; auto shutdown on interlock breach.

3. Detector Requirements:
- Single solid state detector; simultaneous reading of all wavelengths.

4. RF Generator Requirements:
- Solid state generator, 750-1500 W with high power transfer efficiency.

5. Gas Flow Controls:
- All plasma gases must use mass flow controllers meeting specified flow ranges.

6. Sample Introduction System:
- Cassette style vertical torch; double pass glass cyclonic spray chamber + glass concentric nebulizer.

7. Software Requirements:
- 64-bit software; Windows 11 compatible with automated background correction capabilities.

8. Performance Requirements:
- Must meet all EPA ICP OES detection limits using specified sample introduction systems.

9. Additional Requirements:
- New instrument only; refurbished/demo units not acceptable with a one-year warranty.

Period of Performance
Preferred delivery time is within 30 days.

Place of Performance
2744 Rodman Ave. Rock Island, Illinois 61299

Bidder Requirements
This opportunity is a Total Small Business Set-Aside (SBA). Bidders must provide new equipment only; used or remanufactured products will not be accepted. Bidders must also ensure compliance with Buy American Act provisions and provide certifications regarding country of origin.
MANDATORY MINIMUM REQUIREMENTS: Preferred instrument: ICP- OES comparable to the Agilent 5800; vendors may offer any equivalent system that meets the required specifications.1. Spectrometer RequirementsSimultaneous reading ICP OES w/ solid state detector.Bench top design only. Floor mount not acceptable.Echelle polychromator, 253 mm focal length, single detector, single entrance slit. Resolution: less than 0.0065 nm @ 200 nm; less than 0.013 nm @ 327.4 nm.Purged + thermostated optical enclosure; purge 0.7 L/min w/ 3 L/min boost option (less than 189 nm).Pixel level anti blooming detector.Vertical plasma; single piece torch. Optional torches must be available.View equal to or greater than 96% emission spectrum from 167-785 nm.Simultaneous UV + Vis determinations in one measurement.2. ICP System RequirementsBench top system; dual view vertical plasma (synchronous axial/radial).Must run aqueous + organic matrices including oil, MIBK, kerosene, hexanes.Monitors gas pressures/flows, water flows, air pressure, exhaust, plasma stability; auto shutdown on interlock breach.Operates at 15-30 degrees C and 25-80% RH; tolerates 2 degrees C/hr change with no performance loss.Exhaust requirement: equal to or greater than 2.5 m^3/min (88 ft^3/min).Power: equal to or less than 2.9 kVA, 200-240 VAC, 50-60 Hz, equal to or less than 15 amps3. Detector RequirementsSingle solid state detector; simultaneous reading of all wavelengths.Peltier cooled to equal to or less than -40 degrees C; Adaptive Integration capability.Hermetically sealed; no gas purge required.4. RF Generator RequirementsSolid state generator, 750-1500 W, 10 W increments.equal to or greater than 75% power transfer efficiency; typical stability 0.1%.5. Gas Flow ControlsAll plasma gases (coolant, auxiliary, nebulizer, makeup) must use mass flow controllers.Must meet flow ranges + increments listed in Government technical criteria.Integrated, computer controlled oxygen addition for organics.6. Sample Introduction SystemCassette style vertical torch; no manual adjustments required after mounting.Double pass glass cyclonic spray chamber + glass concentric nebulizer.Five channel variable speed peristaltic pump; up to 80 rpm.7. Software Requirements64 bit software; Windows 11 compatible.Automated background + spectral interference correction; automatic IEC recalculation.QC system supporting blanks, spikes, duplicates, and corrective actions.Full raw data saving + complete post run reprocessing.8. Performance RequirementsMust meet all EPA ICP OES detection limits and CRDLs using concentric nebulizer + cyclonic spray chamber.9. Additional RequirementsNew instrument; refurbished/demo units not acceptable.One year warranty (parts, labor, travel).Separate, remotely placeable water chiller.Autosampler holding equal to or greater than 180 sample positions + equal to or greater than 11 standards10. Hydrofluoric Acid Compatibility RequirementSample introduction systems must be or include compatible hardware for hydrofluoric acid use. This shall include, at a minimum, an HF-resistant nebulizer and spray chamber11. Certified Metal Standards Accuracy RequirementThe instrument shall be capable of analyzing within +/- 10% of certified values of metal standards (either chip or atomized metal powder) for typical alloy steel grades, typical stainless steel grades, typical copper alloys, typical cast and wrought aluminum, and titanium alloys with proper preparation for metallurgical elements excluding elements atypical of interrogation by ICP-OES methods such as carbon, nitrogen, hydrogen, and oxygen.

Overview

Response Deadline
Due in 7 Days
Posted
Set Aside
Small Business (SBA)
Place of Performance
2744 Rodman Ave. Rock Island, Illinois Rock Island, IL 61299 USA
Source

Current SBA Size Standard
1000 Employees
Pricing
Likely Fixed Price
Est. Level of Competition
Average
Est. Value Range
Experimental
$85,000 - $160,000 (AI estimate)
Signs of Shaping
The solicitation is open for 7 days, below average for the MICC Fort Drum. 91% of obligations for similar contracts within the Department of the Army were awarded full & open.
On 10/1/26 MICC Fort Drum issued Synopsis Solicitation W911S227U0002 for Inductively Coupled Plasma - Optical Emission Spectrometry due 10/9/26. The opportunity was issued with a Small Business (SBA) set aside with NAICS 334516 (SBA Size Standard 1000 Employees) and PSC 6630.
Primary Contact
Name
Alec Pirolli   Profile
Phone
(520) 692-9834

Documents

Posted documents for Synopsis Solicitation W911S227U0002

Opportunity Assistant


Analyze Opportunity

Generate a Draft

Incumbent or Similar Awards

Contracts Similar to Synopsis Solicitation W911S227U0002

Potential Bidders and Partners

Awardees that have won contracts similar to Synopsis Solicitation W911S227U0002

Similar Active Opportunities

Open contract opportunities similar to Synopsis Solicitation W911S227U0002

Experts for Inductively Coupled Plasma - Optical Emission Spectrometry

Recommended subject matter experts available for hire

Additional Details

Source Agency Hierarchy
DEPT OF DEFENSE > DEPT OF THE ARMY > AMC > ACC > MISSION & INSTALLATION CONTRACTING COMMAND > 419TH CSB
FPDS Organization Code
2100-W911S2
Source Organization Code
500036690
Last Updated
Last Updated By
alec.l.pirolli.civ@army.mil
Archive Date