A24-007 Water Cooled Xray System 76 AMXG Request for the purchase of 4 ea.

Water Cooled Xray System to include the following:

Input Line Requirements:

Automatically adapts to input line voltage100-130Vac, 50/60Hz, 20 Amperes maximum.

200-250Vac, 50/60Hz, 10 Amperes maximum.

10 to 160kV, 0.5mA to 5.0mA (800 watts max.)

Constant potential, end-grounded anode, liquid cooled version.

X-Ray Tube Window: Beryllium 0.8mm (directional) 40 directional.

Radiation Output: 14R/min at 50cm filtered with 0.5 inches (12.7mm).

aluminum at 160kV, 5mA.

Radiation Leakage: Less than 2.0 Roentgens per hour at 1 meter from the X-Ray tube target.

Effective Focal Spot: Standard: 0.7mm x 1.0mm (0.03in x 0.04in) EN125430.5mm x 0.5mm (0.02in x 0.02in) nominal IEC60336.

Ambient Temp: 100% duty cycle @ 120 F (49 C).

Duty Cycle: 100%.

Storage Temp: -30 F to 160 F (-35 C to 71 C).

Anode Cooling: Liquid coolant solution closed loop between X-Ray tube anode and cooling unit.

Liquid Cooling Unit Weight: 54lbs. (15.4kg) approx.

Tube Head Dimensions: No greater than 7.25" Diameter, 30" Length.

Liquid Cooled: 29lbs. (13.15kg) X-Ray Control Unit: X-Ray Control Unit Weight: 26lbs. (11.8kg) approx.

Safety Devices:

Tubehead Pressure Relief Valve

Tubehead Thermal Cut-Out

Tubehead Pressure Gauge

Tubehead Low Pressure Cut-Out @ 25psi (1.72 Bar)

Coolant Flow Sensor (liquid cooled only)

Control Unit Safety Keyswitch

Microcomputer-based Self Diagnostics

Continuous Exposure Parameter Display

Standard Accessories:

Operation manual

Tubehead carrying case

Tubehead Cable- 100' with strain relief

Extra key (1) for Control Unit Safety Lock

Power cable - 25' with strain relief

Coolant hose - twin, 50' with self-sealing terminations

(liquid cooled only)

Cooler power cable

GUI Control Software for the GUI is specifically designed for controlling series systems. As an alternative to the front panel control, the GUI will allow the user to control all necessary functions of the system from a user-friendly windows-based menu. Additionally, the GUI can be used as a diagnostic tool when the system is controlled via the front panel.

Background
The contract is issued by the United States government for the procurement of a Water Cooled X-ray System, specifically identified as A24-007. The goal of this contract is to acquire four units of this specialized medical imaging equipment, which is essential for various applications in healthcare and diagnostics. The agency aims to ensure that the necessary equipment is available to support its mission effectively.

Work Details
The contract requires the delivery of 4 units of the Water Cooled X-ray System with the following specifications:
- Input Line Requirements:
- Automatically adapts to input line voltage 100-130Vac, 50/60Hz, 20 Amperes maximum.
- 200-250Vac, 50/60Hz, 10 Amperes maximum.

- Voltage and Current:
- 10 to 160kV, 0.5mA to 5.0mA (800 watts max.)

- Design: Constant potential, end-grounded anode, liquid cooled version.

- X-Ray Tube Window: Beryllium 0.8mm (directional) at a 40° angle.

- Radiation Output: 14R/min at 50cm filtered with 0.5 inches (12.7mm) aluminum at 160kV, 5mA.

- Radiation Leakage: Less than 2.0 Roentgens per hour at a distance of 1 meter from the X-Ray tube target.

- Effective Focal Spot: Standard dimensions are 0.7mm x 1.0mm (0.03in x 0.04in) and nominal dimensions are EN125430.5mm x 0.5mm (0.02in x 0.02in).

- Ambient Temperature: Operates at a duty cycle of up to 100% at temperatures up to 120°F (49°C).

- Storage Temperature: Ranges from -30°F to 160°F (-35°C to 71°C).

- Anode Cooling: Features a liquid coolant solution in a closed loop between the X-Ray tube anode and cooling unit.

- Weight Specifications:
- Liquid Cooling Unit Weight: Approximately 54lbs (15.4kg).
- Tube Head Dimensions: No greater than a diameter of 7.25 inches and length of 30 inches.
- Liquid Cooled Unit Weight: Approximately 29lbs (13.15kg).
- Control Unit Weight: Approximately 26lbs (11.8kg).

- Safety Devices Include:
- Tubehead Pressure Relief Valve
- Tubehead Thermal Cut-Out
- Tubehead Pressure Gauge
- Tubehead Low Pressure Cut-Out @25psi (1.72 Bar)
- Coolant Flow Sensor
- Control Unit Safety Keyswitch
- Microcomputer-based Self Diagnostics
- Continuous Exposure Parameter Display

- Standard Accessories Provided:
- Operation manual
- Tubehead carrying case
- Tubehead Cable –100’ with strain relief
- Extra key for Control Unit Safety Lock
- Power cable –25’ with strain relief
- Coolant hose –twin,50’ with self-sealing terminations (liquid cooled only)
- Cooler power cable
- GUI Control Software designed for controlling series systems.

Place of Performance
Deliveries will be made to Tinker Air Force Base, Oklahoma.

Overview

Response Deadline
Jan. 21, 2025, 4:00 p.m. EST Past Due
Posted
Jan. 8, 2025, 1:04 p.m. EST
Set Aside
Small Business (SBA)
Place of Performance
Tinker AFB, OK 73145 United States
Source
SAM

Current SBA Size Standard
600 Employees
Pricing
Likely Fixed Price
Est. Level of Competition
Average
Est. Value Range
Experimental
$400,000 - $600,000 (AI estimate)
Signs of Shaping
88% of obligations for similar contracts within the Department of the Air Force were awarded full & open.
On 1/8/25 Air Force Sustainment Center issued Solicitation FA812525Q0027 for A24-007 Water Cooled Xray System due 1/21/25. The opportunity was issued with a Small Business (SBA) set aside with NAICS 334519 (SBA Size Standard 600 Employees) and PSC 6680.
Primary Contact
Name
Sondra Callaway   Profile
Phone
None

Secondary Contact

Name
Darla DeAngelis   Profile
Phone
None

Documents

Posted documents for Solicitation FA812525Q0027

Question & Answer

Opportunity Lifecycle

Procurement notices related to Solicitation FA812525Q0027

Award Notifications

Agency published notification of awards for Solicitation FA812525Q0027

Incumbent or Similar Awards

Contracts Similar to Solicitation FA812525Q0027

Potential Bidders and Partners

Awardees that have won contracts similar to Solicitation FA812525Q0027

Similar Active Opportunities

Open contract opportunities similar to Solicitation FA812525Q0027

Additional Details

Source Agency Hierarchy
DEPT OF DEFENSE > DEPT OF THE AIR FORCE > AFMC > AIR FORCE SUSTAINMENT CENTER > FA8125 AFSC PZIMA
FPDS Organization Code
5700-FA8125
Source Organization Code
500020988
Last Updated
Feb. 5, 2025
Last Updated By
sondra.callaway@us.af.mil
Archive Date
Feb. 5, 2025